Tuesday, July 24, 2012
Future Development of Multislice CT
Tuesday, July 17, 2012
Dual-source CT
Thursday, July 12, 2012
Collimators Component on CT System
- to reduce unnecessary radiation doses to patient
- To ensure good image quality
- Prepatient : the position of collimator between the xray source and the paient. Since x ray photons emitted from the x ray cover a very wide range in z, the prepatient collimator restrict the x ray flux applied to narrow region as it sis applied to the patient. For a single slice CT, it not only reduces dose to the patient. It also defines the slice thickness of t he imaging plane. For multislice CT, however, the slice thickness is defined by detector aperture instead of the collimator. Because nearly 99% of the x ray photons emitted from the xray tube blocked by pre patient collimator , x ray efficiency for CT is poor.
- Postpatient : Typically, two kinds of collimators are used : in plane and cross plane. The in plane collimation (grid)n is used by third generation CT scanner to reject scattered x ray photons. This technique cannot be employed effectively by foruth generation scanners due to a wide acceptance angle. This type sre place in front of the detector, focusing on the x ray source. Since the path of scateered raediation generally deviates from the original x ray photon (primary photon) path, plat block the photon from watering the detector.Cross plane collimation is employed by both third and fourth generation scanner. For 3rd generation scanner, it servers mainly as additional z collimator to improve the slice sensitivity profile of scanner. The slice thickness is determined by combination of umbra and penumbra region. Because of the geomethric limitation, it is often difficult to design a prepatient collimator that provides very thin slices profiles for the scanner . The achieve this objective, additional collimation is employed near the surface of the detector to futher restrict the x ray beam to a narrow slice thickness. The disadvantage to approach is the penalty to the dose efficiently of the scanner because a portion of the x ray photons that pass through the patient is not used
Wednesday, July 11, 2012
Gantry and Slip Ring
The X-ray Tube a Major component of CT Scanner.
Wednesday, June 27, 2012
Four CT Scanner Generations
From historical perspective, four
scanner generations are known in CT. Head scanners, which the patient by translation
and rotation of the measurement system with pencil beam(a) or a small fan
beam(b) and fan beam system. The “3rd Generation” featuring a
rotating detector(c)has clearly outdistance the “4th generation”
,which utilizes stationary detector rings(d)Thursday, June 14, 2012
Filter Component of CT System
Tuesday, June 05, 2012
Stress Testing on CT & Changes in Reading Images Comparing 64-Slice to 256-, 320-Slice CT
Stress Testing upon CT
Min said the actual higher-slice systems may present an advantage in perfusion image resolution compared with 64-slice, because of their capability to image the entire coronary heart at once. Up until now, he explained only 64-slice systems happen to be evaluated for CT perfusion image resolution. The higher-slice systems will probably show CT can be used for high-quality heart perfusion imaging, but studies are needed. These systems also provide additional perfusion applications with regard to liver and mind imaging. This may assist justify the higher price, if shared in between several specialities.
The actual Aquilion One 320-slice scanner may complete an entire physique scan in 3 seconds and a coronary heart scan without desk movement in less than fifty percent a second. Poon said this particular speed and protection will allow CT adenosine stress image resolution, which is primarily restricted to magnetic resonance imaging (MRI) as well as nuclear imaging.
“Nuclear continues to be the standard of treatment (for perfusion imaging), however the temporal spatial resolution isn't good at all,” Poon stated. He added the reduced resolution can skip small perfusion defects which are easily seen around the 320-slice images.
Changes in Studying Images
The minor misalignments in 64-slice cardiac pictures lead to many items, which radiologists and cardiologists consider when reading. Nevertheless, a 320-slice scanner picture eliminates such items. Poon said what you see may be the actual, full coronary heart frozen at a particular time.
“I had to alter my reading routines,” Poon said. “I was so accustomed to reading 64-slice that I simply assumed many things We saw were items. But now what I 'm seeing is a sore; it’s not an artifact anymore. You have to look at the 330 under a different mild. You can’t assume you will find artifacts anymore.”
The actual University of Chi town Medical Center uses 16-, 64- as well as 256-slice systems. Some individuals had exams carried out at different occasions by two or even all three of the methods. Vannier said the radiologists in comparison these images as well as were able to pick out the actual 256-slice system immediately. Just about all motion was freezing and the images didn't have the “blocky” misaligned sewing artifacts, he said. Our prime motion area close to the diaphragm also contributed to movement blur that made an appearance as a thickening of cells, which would have resulted in an improper prognosis. “We started asking ourself if we were phoning things abnormal once they were actually Okay. It was really sobering,” Vannier stated.
Tuesday, May 15, 2012
Cardiac CT With >64 Slices
- Currenly the only commercially available 256 Slice CT Scanner(Philips) has a 128 detector configuration with double sample. Air bearings allow a rapid gantry location time of 0,027s, giving a temporal resolution of 135ms. The unit is larger that typical 64 slice scanner due to the wider detector array and the need with stand the high g-forces generated by rapid gantry rotation.
- Z axis coverage is double(8 cm) that of a typical 64-slice scanner, but still short enough to require at least two gantry rotations to cover the heart completely, images may still be susceptible to step and misregistration artifacts. However , as fewer overlapping slices are required to cover the heart, scan acquisition time reduced.
- The only commercially available 320 detector scanner (Toshiba) has ab ultrawide detector width of 16cm(320 x 0.5mm), which is wide enough to cover the entire heart in a single rotation without the need to move the patient through the gantry. The 350ms gantry rotation time gives a temporal resolution of 175ms using half scan reconstruction . The scanner has several technology advantages
- Low radiation dose :single rotation cardiac acquisition allow complete data set to be obtained in a single prospective pulse. As there is no need for overlap between successive rotations, radiation dose is reduced(typically 1-3 mSv)
- Arrthyhmia management :single rotation cardiac acquisition enables the scanner to wait for an appropriately long R-R interval before the acquisition ‘pulse’. This allows a static, isophasic datasheet in all patients providing ventricular rate is well controlled(heart rate <70 bpm)
- Prospective single-beat functional : ultrawide detector array allows acquisition of functional data from all phases of cardiac cycle using only a single heartbeat
- The wide cone-beam angle (>14o) requires substantial computer processing power to produce artifact -free images, particularity to corrector scatter
- Very wide detectors are substantially heavier than conventional detectors, resulting in higher centrifugal forces during rotation
- As a result , temporal resolution is limited to 175 ms restricting single beat prospective scanning to patient with heath rate<65bpm.
Wednesday, January 18, 2012
How a CT Scan?
Doing a CT scan is the reason CT (computed tomography) scan to scan a clear picture of the various internal organs, including bone, a variety of tissues, and blood vessels. The scans are very useful in the diagnosis of various diseases and conditions, including cancer, heart disease, vascular disease, stroke, reproductive problems, and digestive problems.
Unlike conventional scanning X-ray, CT scan is able to scan the organs in much more detail to allow doctors to diagnose correctly. CT scans are not always conducted in a serious situation and does not necessarily mean the patient is suffering from a serious condition. CT scan is a diagnostic tool regardless of the severity of the patient.
CT scan machine looks like a large tube which is equipped with some sort of table where the patient is lying down. An X-ray scanners at the tube and then scan the patient's body. Every single round of scanning about 1,000 pictures taken of the body. This image is then compiled by a computer to form a two-dimensional cross-sectional images of the body which are then analyzed.
Preparation of CT Scan
CT scans do not require elaborate preparation. Will be very helpful if the patient wear loose clothing, remove the metal accessories such as jewelry or glasses, and drink only clear liquids (water) before the CT scan. Depending on the reason for a CT scan, the patient may be asked to swallow a contrast agent prior to scanning. Make the scan contrast agents have a better resolution to facilitate the analysis. Contrast agents are often given to patients with CT scans that have digestive problems.
Consideration
CT scans usually takes 10 to 45 minutes, or longer in some cases. For this reason, CT scans were performed only when doctors feel the procedure is really necessary. Due to radiation exposure of the body, a CT scan is generally performed at most twice a year. Pregnant women in the first trimester is recommended not to undergo a CT scan or other X-ray procedures. While helping to get better results scanning, contrast agents can cause allergic reactions such as nausea and vomiting. In the event of allergy in patients who underwent more than one CT scan should consult a doctor because of an allergic reaction to contrast agents for future use is generally more severe than the first reaction
Saturday, December 24, 2011
How The difference between PET Scan and CT Scan?
PET and CT scans include the use of high-tech equipment to picture the internal structure of the physique. PET scan can expose changes at the cellular degree, while the CT scan showed a general change in the overall structure of the body organ or tissue. Because of this capability of PET scans may detect the disease at an initial phase that can not be done by normal CT scan.
PET Scan
Dog stands for "positron emission tomography". Scanning (check out) uses a small amount of radioactive material tend to be inserted in the body by shot, inhaled, or swallowed. Dog scan will show where the radioactive materials accumulates to detect issues in blood flow, oxygen usage, and metabolism.
CT Scan
CT means "computerized tomography" (also referred to as CAT scan). CT check out uses X-rays to obtain cross-sectional images from the body. These images can be seen individually (two dimensional) or even combined to produce three-dimensional view.
Perform
PET scans can identify the spread of cancer, heart disease identified, and examined mind disorders, and nervous system.As the CT scan can quickly identify accidents due to accidents and other stress, evaluate the condition of individuals after surgery and radiation treatment, as well as the detection of most cancers and vascular disease.
Impact
PET and CT scan is really a procedure that has been used for decades as well as proven to have minimal negative effects. But still there are some things to consider. Each types of scanning involves the utilization of radiation. CT scans may somewhat increase the risk of cancer improvement. However, this risk is extremely small compared with the benefits of the actual CT scan itself. CT scan ought to be used only when necessary. Expecting or lactating women ought to inform their doctor prior to undergoing PET or CT check out.
Scanning Combination
PET checking can scan combined with CT check out. This combination makes the diagnosis can be created more accurately.
Saturday, December 17, 2011
X-ray Detector on CT Sanner Component
Monday, December 12, 2011
CT Scanner Component: High Voltage Generator
- If cathode voltage becomes positive relative to anode, electron emitted from the hot anode will accelerate toward the cathode, causing the cathode filament to age prematurely and generate undesired x ray photons(located significant off the x ray focal spot)
- High fluctuation of the voltage causes significant difficulties in the calibration and data conditioning for the CT systems. Note that the energy spectrum of the x ray flux is linked closely to the voltage potential across the cathode and anode. When significant variation exist in the power supply voltage, the x ray spectrum changer constantly from view to view. This makes some of correction process, such as beam hardening, nearly impossible. To ensure adequate image quality and prevent premature x ray tube failure, the power supply to the x ray is typically rectified and additional efforts are made to ensure the voltages is as close to a constant as possible ( to simulate a dc power supply). The voltage generator from single phase rectified typically has ripple of 100 %. Ripple is defined as t he peaks to valley voltage over peak voltage