Here is advantages and disadvantages of Cardiac CT With >64 Slices
256 Slice CT
- Currenly the only commercially available 256 Slice CT Scanner(Philips) has a 128 detector configuration with double sample. Air bearings allow a rapid gantry location time of 0,027s, giving a temporal resolution of 135ms. The unit is larger that typical 64 slice scanner due to the wider detector array and the need with stand the high g-forces generated by rapid gantry rotation.
- Z axis coverage is double(8 cm) that of a typical 64-slice scanner, but still short enough to require at least two gantry rotations to cover the heart completely, images may still be susceptible to step and misregistration artifacts. However , as fewer overlapping slices are required to cover the heart, scan acquisition time reduced.
320 Slice CT
- The only commercially available 320 detector scanner (Toshiba) has ab ultrawide detector width of 16cm(320 x 0.5mm), which is wide enough to cover the entire heart in a single rotation without the need to move the patient through the gantry. The 350ms gantry rotation time gives a temporal resolution of 175ms using half scan reconstruction . The scanner has several technology advantages
- Low radiation dose :single rotation cardiac acquisition allow complete data set to be obtained in a single prospective pulse. As there is no need for overlap between successive rotations, radiation dose is reduced(typically 1-3 mSv)
- Arrthyhmia management :single rotation cardiac acquisition enables the scanner to wait for an appropriately long R-R interval before the acquisition ‘pulse’. This allows a static, isophasic datasheet in all patients providing ventricular rate is well controlled(heart rate <70 bpm)
- Prospective single-beat functional : ultrawide detector array allows acquisition of functional data from all phases of cardiac cycle using only a single heartbeat
However , there also several disadvantages
- The wide cone-beam angle (>14o) requires substantial computer processing power to produce artifact -free images, particularity to corrector scatter
- Very wide detectors are substantially heavier than conventional detectors, resulting in higher centrifugal forces during rotation
- As a result , temporal resolution is limited to 175 ms restricting single beat prospective scanning to patient with heath rate<65bpm.
No comments:
Post a Comment